Could I help you?

ASTM E1607-96e1

ASTM E1607-96e1 [ Withdrawn ] Standard Practice for Use of the Alanine-EPR Dosimetry System (Withdrawn 2002)

standard by ASTM International, 05/01/1996

More details

Reduced price!
M00088571
New product
Limited time offer:

$26.10

$58.00

-55%

In stock

- +

Add to compare

 
More info

Full Description

1.1 This practice covers materials description, dosimeter preparation, instrumentation, and procedures for using the alanine-EPR dosimetry system for measuring the absorbed dose in materials irradiated with photons and electrons. The system is based on electron paramagnetic resonance (EPR) spectroscopy of the amino acid alanine. It is classified as a reference standard dosimetry system (see Guide E1261).

1.2 This practice covers alanine-EPR dosimetry systems for dose measurements under the following conditions:

1.2.1 The absorbed dose range is between 1 and 10 Gy.

1.2.2 The absorbed dose rate is up to 10 Gy s -1 for continuous radiation fields and up to 10 Gy s -1 for pulsed radiation fields (1, 2).

1.2.3 The radiation energy range for photons is between 0.1 and 50 MeV and for electrons is between 0.3 and 50 MeV (1, 2).

1.2.4 The irradiation temperature is between -60 and +90

Product Details

Published: 05/01/1996 Number of Pages: 7 File Size: 1 file , 46 KB

 
Custom tab

This is a custom block edited from admin panel.You can insert any content here.

30 other products in the same category:

Compare 0

No products

To be determined Shipping
$0.00 Total

Check out